Probe CP-2

Login to see price

Not Available For Sale


  • Brand
  • Choose Your Handle

This combination does not exist.

Brand: ErgoDenta

ErgoDenta
ErgoDenta

Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.

Explore the ErgoDenta range →

Website Product Category :
Diagnostics , Periodontal Probes


The Probe CP-2 is a periodontal probe from ErgoDenta™'s ErgoTip™ interchangeable system, used for measuring pocket depth and assessing periodontal health. Screw-on tip fits all ErgoTip™ handles.

Type Periodontal Probe
Use Pocket measurement
Material Stainless Steel
i

Clinical Guide: Choosing the Right Periodontal Probe

New to probes or not sure which tip, marking style, or handle is right for your workflow? Our detailed buying & usage guide walks you through CP 2, CP 15, WHO, Williams, Marquis, Nabers and Goldman-Fox probes, explains the colour-coded depth markings, compares the ErgoDenta Standard, ErgoLite and ErgoTip systems, and helps you pick the handle weight & material that best suits your hand.

Read the full Periodontal Probes guide →

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.