Spoon Excavator, 1.7mm | ErgoDenta
Spoon Excavator, 1.5mm | ErgoDenta
Spoon Excavator, 1.5mm | ErgoDenta
Login to see price
Spoon Excavator, 2.0mm | ErgoDenta
Spoon Excavator, 2.0mm | ErgoDenta
Login to see price

Spoon Excavator, 1.7mm | ErgoDenta

2814 SP
https://ergodenta.dk/web/image/product.template/11298/image_1920?unique=c44b698

Spoon Excavator, 1.7mm

Login to see price

Not Available For Sale


  • Brand
  • Choose Your Handle

This combination does not exist.

Brand: ErgoDenta

ErgoDenta
ErgoDenta

Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.

Explore the ErgoDenta range →

Website Product Category :
Restoration , Excavators


The Spoon Excavator, 1.7mm is designed for precise removal of carious dentin and soft tissue from cavity preparations. Its sharp, spoon-shaped working end provides excellent tactile feedback and controlled excavation, making it ideal for conservative restorative procedures. The ergonomic handle ensures comfortable grip during extended use.

Key Features

  • 1.7mm spoon-shaped working end for precise excavation
  • Sharp cutting edges for efficient caries removal
  • Ergonomic handle design for comfortable grip
  • Excellent tactile control during cavity preparation
  • Suitable for both anterior and posterior teeth
  • High-quality stainless steel construction

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.