Microscopic HD Mouth Mirror, UltraThin 0.7mm, Black | ErgoDenta
Microscopic HD Mouth Mirror, Regular 1.5mm, Black | ErgoDenta
Microscopic HD Mouth Mirror, Regular 1.5mm, Black | ErgoDenta
Login to see price
Microscopic HD Mouth Mirror, Regular 1.5mm, Black | ErgoDenta
Microscopic HD Mouth Mirror, Regular 1.5mm, Black | ErgoDenta
Login to see price
Best Seller

Microscopic HD Mouth Mirror, UltraThin 0.7mm, Black | ErgoDenta

7093U
https://ergodenta.dk/web/image/product.template/11456/image_1920?unique=cd125c3

Microscopic HD mouth mirror 7mm, ultra-thin 0.7 mm, anti-glare coating for endodontics.

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoDenta

ErgoDenta
ErgoDenta

Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.

Explore the ErgoDenta range →

Website Product Category :
Mirrors , HD Micro Mirrors


The Microscopic HD Mouth Mirror, UltraThin 0.7mm, Black is a microscope-grade mouth mirror for endodontics. A 36-layer anti-glare coating gives a crisp, reflection-free image in the smallest spaces; supplied with an ErgoLite mirror handle.

  • 36-layer anti-glare coating for a clear, reflection-free image
  • Ultra-thin 0.7 mm glass — maximises access and visibility
  • Built for microscope-assisted endodontics
  • Supplied with an ErgoLite mirror handle
Working size7mm
Coating36-layer anti-glare
Thicknessultra-thin 0.7 mm
SterilisationFully autoclavable
Brand lineErgoDenta

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.