Perforated Impression Tray – Lower Jaw – Size XS
4501-1
Perforated stainless steel tray for lower-jaw dental impressions.
Website Product Category :
Impression TraysTags :
The Perforated Impression Tray – Lower Jaw – Size XS is made of stainless steel and is intended for routine dental impression procedures on the lower arch. Its perforated body lets the impression material lock through the openings, and the Rim-Lock rim helps retain the set material on removal. The tray is reusable and fully autoclavable, making it suitable for repeated use in busy clinics. A stable stainless steel structure resists flexing during the impression, and the handle gives a firm grip when seating and lifting the tray. Size XS suits extra small lower-jaw arches encountered in daily practice.
Key features
- Stainless steel construction
- Reusable and fully autoclavable
- Rim-Lock system for secure impression retention
- Perforated design for mechanical material retention
- Lower-jaw configuration
- Size XS (Extra Small)
- Suitable for routine adult dental impression procedures
Specifications
- Article #: 4501-1
- Size: XS (Extra Small)
- Jaw: Lower
- Design: Perforated
- Retention: Rim-Lock system
- Material: Stainless steel
- Sterilisation: Fully autoclavable
Reprocessing Instructions
Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.