Basic Diagnostic Kit
9908 SP
Basic Diagnostic Kit · 5 instruments + sterilisation cassette.
Basic Diagnostic Kit — a practical starter set for everyday dental examinations. Supplied in a sterilisation cassette for organised storage and efficient sterilisation.
What's inside
- 1401 — ErgoX® Mirror handle, Peach Blush
- 1104 — Explorer 5-9
- 1309 — Explorer 23 + WHO Probe (double-ended)
- 2802 — Round Excavator, 1.5 mm
- 3627-1A — College Tweezer, Serrated, 15 cm
5 instruments + sterilisation cassette
Specifications
- Article #: —
- Contents: 5 instruments + sterilisation cassette
- Material: Surgical stainless steel
- Sterilization: Autoclavable, all standard cycles
Reprocessing Instructions
Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.