Luxating Elevators Kit Gold — 8 pcs, Standard Handle
Luxating Elevators Kit — 8 pcs, Standard Handle
Luxating Elevators Kit — 8 pcs, Standard Handle
Login to see price
Luxating Elevators Kit Gold — 8 pcs, Classic Handle
Luxating Elevators Kit Gold — 8 pcs, Classic Handle
Login to see price

Luxating Elevators Kit Gold — 8 pcs, Standard Handle

6001G
https://ergodenta.dk/web/image/product.template/22425/image_1920?unique=2d5b1f4

Luxating Elevators Kit Gold — 8 pcs, Standard Handle — 8-piece luxating elevator set (standard ergonomic handle with gold-coated tips for high visual contrast and reduced friction). Four straight (2.5/3/4/5 mm) and four curved (2.5/3/4/5 mm) luxators for atraumatic extraction, supplied tray-organised.

Login to see price

Not Available For Sale


This combination does not exist.


Website Product Category :
Surgery , Luxating Elevators
Tags :


The Luxating Elevators Kit Gold — 8 pcs, Standard Handle is a ready-to-use 8-piece luxating elevator set from ErgoDenta, mounted on the standard ergonomic handle with gold-coated tips for high visual contrast and reduced friction. Luxators cut the periodontal ligament and gently expand the socket for atraumatic extraction — preserving bone for implant placement. Supplied tray-organised for fast set-up, sterilisation and reprocessing.

Finish: Gold (TiN-coated tips). Handle: standard ergonomic handle.

What's included (8 pcs):

  • Straight 2.5mm Gold — 6001-1G
  • Straight 3mm Gold — 6001-2G
  • Straight 4mm Gold — 6001-3G
  • Straight 5mm Gold — 6001-4G
  • Curved 2.5mm Gold — 6001-5G
  • Curved 3mm Gold — 6001-6G
  • Curved 4mm Gold — 6001-7G
  • Curved 5mm Gold — 6001-8G

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.

You might also need

Explore more of the ErgoDenta range.

Products you view will appear here.