Universal Upper Roots Extracting Forceps | Standard Extraction Forcep | ErgoDenta
Changeable Part for Posterior Band Removing Plier, Plastic Pad | ErgoDenta
Changeable Part for Posterior Band Removing Plier, Plastic Pad | ErgoDenta
Login to see price
Universal-Lower Roots | Standard Extraction Forcep | ErgoDenta
Universal-Lower Roots | Standard Extraction Forcep | ErgoDenta
Login to see price

Universal Upper Roots Extracting Forceps | Standard Extraction Forcep | ErgoDenta

5005-OK
https://ergodenta.dk/web/image/product.template/11360/image_1920?unique=1201a46

Universal Upper Roots Extracting Forceps

Login to see price

Not Available For Sale


  • Brand
  • Choose Your Handle

This combination does not exist.

Brand: ErgoDenta
Choose Your Handle: English Pattern

ErgoDenta
ErgoDenta

Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.

Explore the ErgoDenta range →

Website Product Category :
Extraction Forceps


A universal forceps designed for extracting various types of upper roots. The versatile beak design adapts to different root anatomies, reducing the need for multiple instruments. Ideal for surgical extractions and removal of broken root fragments in the upper jaw.

Key Features:

  • Universal design
  • Adapts to different anatomies
  • Reduces need for multiple instruments
  • Designed for root extraction
  • Fine beaks for precise access

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.