Perforated Impression Tray Set – 5 Upper & 5 Lower
8, 180x136x26mm, Purple | ErgoTray
8, 180x136x26mm, Purple | ErgoTray
Login to see price
Non-Perforated Impression Tray Set – 5 Upper & 5 Lower
Non-Perforated Impression Tray Set – 5 Upper & 5 Lower
Login to see price

Perforated Impression Tray Set – 5 Upper & 5 Lower

4521
https://ergodenta.dk/web/image/product.template/22474/image_1920?unique=0ff9093

Set of ten perforated stainless steel impression trays.

Login to see price

Not Available For Sale


This combination does not exist.



The Perforated Impression Tray Set is a complete range of ten reusable stainless steel trays — five upper-jaw and five lower-jaw — covering sizes Extra Small, Small, Medium, Large and Extra Large. The perforated body of each tray helps retain the impression material through the openings, while the Rim-Lock rim keeps the set material securely seated during removal. Every tray in the set is fully autoclavable and built for everyday clinical use. Stainless steel construction keeps each tray dimensionally stable through repeated sterilisation, and the handle on each tray allows clean positioning and release. Supplied as a full size range to cover the most common adult dental impression cases.

Key features

  • Stainless steel construction
  • Reusable and fully autoclavable
  • Rim-Lock system for secure impression retention
  • Perforated design for mechanical material retention
  • Ten trays: five upper-jaw and five lower-jaw
  • Full size range: XS, S, M, L, XL
  • Covers routine adult impression cases

Specifications

  • Article #: 4521
  • Contents: 5 upper + 5 lower (XS, S, M, L, XL)
  • Design: Perforated
  • Retention: Rim-Lock system
  • Material: Stainless steel
  • Sterilisation: Fully autoclavable

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.