Mayo Hegar Needle Holder, Straight

3212-1S
https://ergodenta.dk/web/image/product.template/21444/image_1920?unique=fdd135d

Mayo Hegar Needle Holder, Straight needle holder for grasping and driving suture.

Login to see price

Not Available For Sale


This combination does not exist.



The Mayo Hegar Needle Holder, Straight is a needle holder for grasping and driving suture needles precisely during oral surgery.

  • Secure jaws grip the needle without slipping
  • Locking ratchet for controlled suturing
  • Reusable and fully autoclavable
TypeNeedle holder
Length
SterilisationFully autoclavable
SuppliedNon-sterile
Brand lineErgoDenta

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.