ErgoX® Plus - ErgoTip Handle, Magenta Blue | ErgoTip
ErgoX® Plus - ErgoTip Handle, Lemon Green | ErgoTip
ErgoX® Plus - ErgoTip Handle, Lemon Green | ErgoTip
Login to see price
ErgoX® Plus - ErgoTip Handle, Citrus Orange | ErgoTip
ErgoX® Plus - ErgoTip Handle, Citrus Orange | ErgoTip
Login to see price

ErgoX® Plus - ErgoTip Handle, Magenta Blue | ErgoTip

1451-XP
https://ergodenta.dk/web/image/product.template/21965/image_1920?unique=c472abe

ErgoX® Plus Handle, ErgoTip™, Magenta Blue

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoTip

ErgoTip
ErgoTip

A smarter system — replaceable working tips on a single ergonomic handle. Less waste, lower cost per instrument, and a handle that stays with the clinician.

Explore the ErgoTip range →

Website Product Category :
ErgoTip Handles


The ErgoX® Plus is a double-ended resin handle from ErgoDenta™’s ErgoTip™ interchangeable system. The wider Ø13.6mm palm-grip diameter distributes pressure more evenly, reducing muscle tension. Available in 8 colours. Compatible with all ErgoTip™ screw-on tips. Autoclavable. Colour: Magenta Blue.

Type Double-End Resin Handle
Diameter Ø13.6mm
System ErgoTip™ Screw-on

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.