ANGLE Wire Bending Pliers, ◯ Max. Ø 0,7mm /.028” | ErgoDenta
Adams Plier, max Ø 0.4mm, Lap joint | ErgoDenta
Adams Plier, max Ø 0.4mm, Lap joint | ErgoDenta
Login to see price
Angle Plier, max Ø 0.7mm, Lap Joint | ErgoDenta
Angle Plier, max Ø 0.7mm, Lap Joint | ErgoDenta
Login to see price

ANGLE Wire Bending Pliers, ◯ Max. Ø 0,7mm /.028” | ErgoDenta

4312
https://ergodenta.dk/web/image/product.template/11010/image_1920?unique=64fd2c1

ANGLE Wire Bending Pliers, 12.5cm

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoDenta

ErgoDenta
ErgoDenta

Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.

Explore the ErgoDenta range →

Website Product Category :
Wire Bending Pliers


The Angle Pliers, max Ø 0.7mm, Box Joint, 12.5cm provides organized storage and easy access to dental instruments and materials. Its durable construction withstands repeated sterilization cycles, while the thoughtful design ensures efficient organization. This holder improves workflow efficiency and maintains a clean, organized clinical environment.

Key Features

  • Organized storage for instruments and materials
  • Durable, autoclavable construction
  • Easy access during procedures
  • Improves workflow efficiency
  • Compatible with standard sterilization equipment
  • Professional-grade quality

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.