Syringe holder | ErgoDenta
Pen Model Intraligamental Syringe, Angled, 1.8 ml, 0.07 ml/ click, with Metric Thread | ErgoDenta
Pen Model Intraligamental Syringe, Angled, 1.8 ml, 0.07 ml/ click, with Metric Thread | ErgoDenta
Login to see price
Sean-Miller Retractor, Sharp | ErgoDenta
Sean-Miller Retractor, Sharp | ErgoDenta
Login to see price

Syringe holder | ErgoDenta

3999
https://ergodenta.dk/web/image/product.template/10960/image_1920?unique=aaddf47

Syringe holder dental anaesthetic syringe for cartridge local anaesthetic.

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoDenta

ErgoDenta
ErgoDenta

Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.

Explore the ErgoDenta range →

Website Product Category :
Other Syringes


The Syringe holder is a dental anaesthetic syringe for delivering local anaesthetic from a standard cartridge.

  • Takes standard anaesthetic cartridges
  • Smooth plunger action
  • Reusable and fully autoclavable
TypeAnaesthetic syringe
Type detailOther Syringes
SterilisationFully autoclavable
SuppliedNon-sterile
Brand lineErgoDenta

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.