Color-coded sterilisation cassettes and instrument trays. Different sizes and colours organise instruments by procedure; perforated design supports efficient autoclave reprocessing.

Explore the ErgoTray range →

Website Product Category :
ErgoTrays


The ErgoVator tray keeps your ErgoLite elevators organised and protected. One tray is included with every ErgoLite elevator. Fully autoclavable and reusable. Colour: Black.

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.