5, 180x86x26mm, Purple | ErgoTray
⚠️Note: We are in the process of updating our website with new product images and information. Some images may be temporarily unavailable. Thank you for your patience!
Classic Surgery Kit
Classic Surgery Kit
Login to see price
8, 180x136x26mm, Purple | ErgoTray
8, 180x136x26mm, Purple | ErgoTray
Login to see price

5, 180x86x26mm, Purple | ErgoTray

5505E
https://ergodenta.dk/web/image/product.template/22446/image_1920?unique=0fde39d

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoTray

ErgoTray
ErgoTray

Keep your instruments organized, protected, and ready to use. ErgoTray is a sterilization cassette and instrument tray system designed to make instrument handling and reprocessing more efficient in daily practice.

ErgoTray cassettes come in different sizes and colors, helping you organize instruments by procedure and identify sets quickly. Perforated designs support effective sterilization, while secure closures and protective interiors help prevent instrument damage during cleaning, sterilization, and storage.

Made from durable materials that withstand repeated autoclaving, ErgoTray systems are built for reliable everyday use. They support smooth instrument workflows, better organization, and high standards of infection control in modern dental clinics.


Website Product Category :
ErgoTrays


The ErgoTray 5 is an autoclavable silicone instrument tray from ErgoDenta™ for organizing up to 5 dental instruments. Flexible, heat-resistant, and easy to clean. 180×86×26mm.

Type Silicone Tray
Use Instrument organization
Size 180×86×26mm

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.