ErgoVator Tray 1, Black
ErgoTray 8, Yellow
ErgoTray 8, Yellow
Login to see price
ErgoVator Tray 1, Grey
ErgoVator Tray 1, Grey
Login to see price

ErgoVator Tray 1, Black

5502K
https://ergodenta.dk/web/image/product.template/11576/image_1920?unique=5f73a27

ErgoVator elevator tray · autoclavable

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoTray

ErgoTray
ErgoTray

Color-coded sterilisation cassettes and instrument trays. Different sizes and colours organise instruments by procedure; perforated design supports efficient autoclave reprocessing.

Explore the ErgoTray range →

Website Product Category :
ErgoTrays


The ErgoVator tray keeps your ErgoLite elevators organised and protected. One tray is included with every ErgoLite elevator. Fully autoclavable and reusable. Colour: Black.

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.