Mirror handle, Double End, Pure Black | ErgoX®
Cone Socket Mirror Handle, 1415C | ErgoX®
Cone Socket Mirror Handle, 1415C | ErgoX®
Login to see price
Front-Surface Mouth Mirror, No. 3 | ErgoDenta
Front-Surface Mouth Mirror, No. 3 | ErgoDenta
Login to see price

Mirror handle, Double End, Pure Black | ErgoX®

1715 X
https://ergodenta.dk/web/image/product.template/22076/image_1920?unique=51d20ab

ErgoX® · ErgoX®.

Login to see price

Not Available For Sale


This combination does not exist.



ErgoX® , Mirror handle, Double End, Pure Black. Designed by ErgoDenta for predictable clinical handling. ErgoX® handle.

Key features

  • Mounted on the ErgoX® handle line
  • Surgical stainless steel — autoclavable

Specifications

  • Article #: 1715 X
  • Handle line: ErgoX®
  • Handle size: Ø11.5mm
  • Handle color / TC: Pure Black
  • Category: ErgoX®, Mirror handles
  • Material: Surgical stainless steel
  • Sterilization: Autoclavable, all standard cycles

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.