Micro Surgical Tweezer, Straight, 18cm, Black
Micro Surgical Tweezer, Straight, 1-2, 15cm, Black
Micro Surgical Tweezer, Straight, 1-2, 15cm, Black
Login to see price
Micro Surgical Tweezer, Curved, 1-2, 18cm, Black
Micro Surgical Tweezer, Curved, 1-2, 18cm, Black
Login to see price

Micro Surgical Tweezer, Straight, 18cm, Black

3802-3
https://ergodenta.dk/web/image/product.template/22262/image_1920?unique=d1e5986

Micro Surgical Tweezer, Straight, 18cm, Black dressing tweezer for handling cotton and.

The 3802-3 Micro-Surgical Tweezer (straight, 18 cm, black) is a fine-tipped instrument for precise tissue handling in microsurgical and aesthetic procedures.

  • Fine tips for micro-surgical precision
  • 18 cm length with black non-glare finish
  • Stainless steel, fully autoclavable

Designed in Denmark by ErgoDenta — fully autoclavable and MDR/FDA-compliant. The Nordic Selection.

Login to see price

Not Available For Sale


This combination does not exist.


Website Product Category :
Black Endo , Black Tweezers , Endodontics


The Micro Surgical Tweezer, Straight, 18cm, Black is a cotton/dressing tweezer for handling cotton pellets, paper points and small materials.

  • Grips cotton pellets, paper points and dressings
  • Serrated tips for a secure hold
  • Reusable and fully autoclavable
TypeCotton/dressing tweezer
SterilisationFully autoclavable
SuppliedNon-sterile
Brand lineErgoDenta

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.