Micro Surgery Kit – 12 pcs (with Sterilization Tray)
9914
Micro Surgery Kit – 12 pcs (with Sterilization Tray) · 12 instruments + sterilisation cassette.
Website Product Category :
SurgeryMicro Surgery Kit — twelve essential instruments for micro-surgical and periodontal-plastic procedures. Organised in a sterilisation cassette for a complete chairside set.
What's inside
- 3008-4D — La Garange Scissor, Double Curved, 11.5 cm, Black, Super-cut, TC
- 5664 — Scalpel Handles No:3
- 1450 SP — ErgoSteel Plus Mirror handle
- 3122-2S — Barraquer / Gingiva Scissor, Straight, 15 cm, TC
- 1504F — Front-Surface Mouth Mirror No. 4, 22 mm
- 7151 SS — Buser, 4 mm–4 mm
- 7106 SP — Lucas 87, 3.5 mm
- 7157 SS — Zucchelli, 3 mm–3 mm
- 7156 SP — Nordberg Molt, 4 mm–7 mm
- 3332-2S — Castroviejo Micro Needle Holder, Straight, 18 cm, TC
- 3802-1G — Micro Surgical Tweezer, Straight, 1–2, 18 cm
- 5513B — Removable-lid sterilisation cassette for 20
12 instruments + sterilisation cassette
Specifications
- Article #: 9914
- Contents: 12 instruments + sterilisation cassette
- Material: Surgical stainless steel
- Sterilization: Autoclavable, all standard cycles
Reprocessing Instructions
Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.