Best Seller
Scalpel Handles No:3 | ErgoDenta
5665
Scalpel Handles No:3
| Brand: ErgoDenta |
ErgoDenta
Designed in Denmark — ergonomic dental hand instruments built for comfortable, efficient daily clinical work across diagnostics, restoration, periodontics, surgery and sterilisation.
Explore the ErgoDenta range →Website Product Category :
Scalpel HandlesTags :
The Scalpel Handles No:3 provides secure blade attachment and precise control during surgical procedures. Its ergonomic design ensures comfortable grip and optimal blade positioning, while the durable construction withstands repeated sterilization. This handle is compatible with standard surgical blades.
Key Features
- Secure blade attachment mechanism
- Ergonomic design for precise control
- Comfortable grip during surgical procedures
- Durable, autoclavable construction
- Compatible with standard surgical blades
- Professional-grade quality
Reprocessing Instructions
Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.