Scalpel Handles No:3
⚠️Note: We are in the process of updating our website with new product images and information. Some images may be temporarily unavailable. Thank you for your patience!
Scaler Tray
Scaler Tray
Login to see price
Scalpel Handles No:3 180°
Scalpel Handles No:3 180°
Login to see price
Best Seller

Scalpel Handles No:3

5665
https://ergodenta.dk/web/image/product.template/11635/image_1920?unique=6eb3640

Scalpel Handles No:3

Login to see price

Not Available For Sale


  • Brand

This combination does not exist.

Brand: ErgoDenta

ErgoDenta
ErgoDenta

ErgoDenta offers ergonomic dental instruments made for comfortable, efficient daily work in the clinic. Our instruments are designed for dentists, dental hygienists, and specialists who need reliable tools they can trust.

The range includes diagnostic instruments, scalers and curettes, restorative tools, surgical instruments, extraction forceps, elevators, endodontic instruments, orthodontic pliers, and sterilization cassettes. Each product focuses on good grip, reduced hand fatigue, and precise clinical performance.

ErgoDenta instruments are built for everyday dental procedures and meet professional sterilization standards. Whether you work in general dentistry, surgery, periodontics, or hygiene, ErgoDenta helps you work with better comfort, control, and confidence.


Website Product Category :
Scalpel Handles


The Scalpel Handles No:3 provides secure blade attachment and precise control during surgical procedures. Its ergonomic design ensures comfortable grip and optimal blade positioning, while the durable construction withstands repeated sterilization. This handle is compatible with standard surgical blades.

Key Features

  • Secure blade attachment mechanism
  • Ergonomic design for precise control
  • Comfortable grip during surgical procedures
  • Durable, autoclavable construction
  • Compatible with standard surgical blades
  • Professional-grade quality

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.