Standard Extraction Forceps Kit, 10pcs
⚠️Note: We are in the process of updating our website with new product images and information. Some images may be temporarily unavailable. Thank you for your patience!
Universal Plier, TC
Universal Plier, TC
Login to see price
Child Extraction Forceps Kit, 7pcs
Child Extraction Forceps Kit, 7pcs
Login to see price

Standard Extraction Forceps Kit, 10pcs

5005
https://ergodenta.dk/web/image/product.template/11027/image_1920?unique=9138df6

Standard Extraction Forceps Kit, 10pcs

Login to see price

Not Available For Sale


  • Brand
  • Choose Your Handle

This combination does not exist.

Brand: ErgoDenta
Choose Your Handle: English Pattern

ErgoDenta
ErgoDenta

ErgoDenta offers ergonomic dental instruments made for comfortable, efficient daily work in the clinic. Our instruments are designed for dentists, dental hygienists, and specialists who need reliable tools they can trust.

The range includes diagnostic instruments, scalers and curettes, restorative tools, surgical instruments, extraction forceps, elevators, endodontic instruments, orthodontic pliers, and sterilization cassettes. Each product focuses on good grip, reduced hand fatigue, and precise clinical performance.

ErgoDenta instruments are built for everyday dental procedures and meet professional sterilization standards. Whether you work in general dentistry, surgery, periodontics, or hygiene, ErgoDenta helps you work with better comfort, control, and confidence.


Website Product Category :
Extraction Forceps


A complete set of 10 standard extraction forceps covering the most common extraction procedures. This kit includes forceps for upper and lower incisors, premolars, molars, and roots. Ideal for general dental practices and provides all essential instruments for routine extractions.

Key Features:

  • Complete 10-piece set
  • Covers all common extractions
  • Durable construction
  • Autoclavable and reusable
  • Professional quality

Reprocessing Instructions

Important: ErgoDenta instruments are supplied NON-STERILE and must be cleaned, disinfected, and sterilized before first use and after each subsequent use.
Step 1
Pre-Cleaning
Rinse under cold running water immediately after use. Remove visible debris with a soft brush. Keep instruments moist until processing.
Step 2
Automated Cleaning
Washer-disinfector: pre-rinse cold, enzymatic wash 55°C/10 min, thermal disinfect 90°C/5 min, dry.
Step 3
Inspect & Maintain
Check for residual soil, corrosion, or damage. Lubricate hinged instruments with medical-grade lubricant.
Step 4
Package
Place in ISO 11607-compliant sterilization pouches or wraps. Arrange to prevent tip contact. Seal properly.
Step 5
Sterilize
Autoclave at 134°C (273°F) for min. 18 minutes, pre-vacuum cycle. Validate with chemical and biological indicators.
Step 6
Store
Store in a clean, dry, enclosed area. Verify packaging integrity before use. Re-sterilize if compromised.